TECHNICAL SPECIFICATIONS$ Generation: DC to AC conversion. Method: 100 Hz vibrations using pure sinusoidal waveform Voltage Display Range: 0 – 50V Test Method Hand held vibrator with mute switch Data Display Instrument: Digital voltmeter Standard Accessories Vibrator- 1No. Hardware unit-1No. Power Cord- 1No. Software CD-1No. Physical Characteristics Instrument Dimensions: 290 mm x 200 mm x 110 mm Approximate Instrument Weight: 3kg Power Requirements Voltage 230 VAC ± 10% Frequency 50 Hz, Power Consumption 1.0 A maximum Minimum PC Requirements OS : Windows* 7 / 8 / 10 CPU : Celeron* 1.8 GHZ / i3 / i5 or better processor HDD : 500GB RAM : 2GB Printer : Any Windows* Compatible Inkjet / Laser Jet Printer | ENVIRONMENT REQUIREMENTS Operating and storage Temp. 15oC to 45oC Ambient Relative Humidity 15% to 90%, non- condensing Operating and storage Pressure/Altitude 523 mm Hg to 760 mm Hg RESULTS Vibration Perception Threshold in Volts for both plantar feet Right Foot: Toe First metatarsal head Third metatarsal head Fifth metatarsal head Instep Heel Left Foot: Toe First metatarsal head Third metatarsal head Fifth metatarsal head Instep Heel Left Foot: |